The application of a new geometry correction function for the calibration of neutron spherical measuring devices using large volume neutron sources
Creators
- 1. Atomic Energy Organization of Iran, Teheran (Iran, Islamic Republic of)
Description
A new geometry correction function for the calibration of spherical dose equivalent survey meters, using large volume neutron sources, has been developed and is presented. This is a modification of the geometry factor formalism developed by Hunt for the correction of non-parallelism of the field which is usually produced by a small neutron source near a spherical detector. This function is used to obtain a functional distance parameter specifying the source virtual centre and the geometry correction factor. The consistency of this correction factor and the source virtual centre, derived in this study, have been investigated by analysing the data of close and far distances using two calibration models applicable in these ranges of distances. (author)
Additional details
Publishing Information
- Journal Title
- Radiation Protection Dosimetry
- Journal Volume
- 44
- Journal Issue
- 1-4
- Journal Page Range
- p. 121-124.
- ISSN
- 0144-8420
- CODEN
- RPDODE
Conference
- Title
- 7. Symposium on neutron dosimetry.
- Dates
- 14-18 Oct 1991.
- Place
- Berlin (Germany).
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- United Kingdom
- INIS RN
- 24038341
- Subject category
- S07: ISOTOPES AND RADIATION SOURCES;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- CALIBRATION; CORRECTIONS; GEOMETRY; MATHEMATICAL MODELS; NEUTRON DOSIMETRY; SPHERES; SURVEY MONITORS
- Descriptors DEC
- DOSIMETRY; MATHEMATICS; MEASURING INSTRUMENTS; MONITORS; RADIATION MONITORS