Inorganic scintillation detectors for 192Ir brachytherapy
Creators
- 1. Department of Radiation Physics, The University of Texas MD Anderson Cancer Center, Houston, TX (United States)
Description
Many brachytherapy (BT) errors could be detected with real-time in vivo dosimetry technology. Inorganic scintillation detectors (ISDs) have demonstrated promising capabilities for BT, because some ISD materials can generate scintillation signals large enough that (a) the background signal emitted in the fiber-optic cable (stem signal) is insignificant, and (b) small detector volumes can be used to avoid volume averaging effects in steep dose gradients near BT sources. We investigated the characteristics of five ISD materials to identify one that is appropriate for BT. ISDs consisting of a 0.26 to 1.0 mm3 volume of ruby (Al2O3:Cr), a mixture of Y2O3:Eu and YVO4:Eu, ZnSe:O, or CsI:Tl coupled to a fiber-optic cable were irradiated in a water-equivalent phantom using a high-dose-rate 192Ir BT source. Detectors based on plastic scintillators BCF-12 and BCF-60 (0.8 mm3 volume) were used as a reference. Measurements demonstrated that the ruby, Y2O3:Eu+YVO4:Eu, ZnSe:O, and CsI:Tl ISDs emitted scintillation signals that were up to 19, 19, 250, and 880 times greater, respectively, than that of the BCF-12 detector. While the total signals of the plastic scintillation detectors were dominated by the stem signal for source positions 0.5 cm from the fiber-optic cable and >3.5 cm from the scintillator volume, the stem signal for the ruby and Y2O3:Eu+YVO4:Eu ISDs were <1% of the total signal for source positions <3.4 and <4.4 cm from the scintillator, respectively, and <0.7% and <0.5% for the ZnSe:O and CsI:Tl ISDs, respectively, for positions ⩽8.0 cm. In contrast to the other ISDs, the Y2O3:Eu+YVO4:Eu ISD exhibited unstable scintillation and significant afterglow. All ISDs exhibited significant energy dependence, i.e. their dose response to distance-dependent 192Ir energy spectra differed significantly from the absorbed dose in water. Provided that energy dependence is accounted for, ZnSe:O ISDs are promising for use in error detection and patient safety monitoring during BT. (paper)
Availability note (English)
Available from http://dx.doi.org/10.1088/1361-6560/ab421fAdditional details
Identifiers
Publishing Information
- Journal Title
- Physics in Medicine and Biology
- Journal Volume
- 64
- Journal Issue
- 22
- Journal Page Range
- [20 p.]
- ISSN
- 0031-9155
- CODEN
- PHMBA7
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 52004130
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ABSORBED RADIATION DOSES; ALUMINIUM OXIDES; BRACHYTHERAPY; CABLES; ENERGY DEPENDENCE; ENERGY SPECTRA; FIBER OPTICS; IN VIVO; IRIDIUM 192; PLASTIC SCINTILLATION DETECTORS; PLASTIC SCINTILLATORS; SIGNALS; YTTRIUM OXIDES
- Descriptors DEC
- ALUMINIUM COMPOUNDS; BETA DECAY RADIOISOTOPES; BETA-MINUS DECAY RADIOISOTOPES; CHALCOGENIDES; DAYS LIVING RADIOISOTOPES; DOSES; ELECTRON CAPTURE RADIOISOTOPES; HEAVY NUCLEI; INTERNAL CONVERSION RADIOISOTOPES; IRIDIUM ISOTOPES; ISOMERIC TRANSITION ISOTOPES; ISOTOPES; MEASURING INSTRUMENTS; MEDICINE; MINUTES LIVING RADIOISOTOPES; NUCLEAR MEDICINE; NUCLEI; ODD-ODD NUCLEI; OPTICS; OXIDES; OXYGEN COMPOUNDS; PHOSPHORS; RADIATION DETECTORS; RADIATION DOSES; RADIOISOTOPES; RADIOLOGY; RADIOTHERAPY; SCINTILLATION COUNTERS; SOLID SCINTILLATION DETECTORS; SPECTRA; THERAPY; TRANSITION ELEMENT COMPOUNDS; YEARS LIVING RADIOISOTOPES; YTTRIUM COMPOUNDS