Published July 1988 | Version v1
Journal article

Effect of gold shielding on the dosimetry of an 125I seed at close range

  • 1. Medical Physics Division, Department of Radiation Oncology, Tufts--New England Medical Center, Boston, Massachusetts 02111

Description

Radioactive 125I plaques with gold shields have been used for the treatment of choroidal melanoma. The effect of the gold shield on the dosimetry of a single 125I seed (model 6711) has been investigated in this paper. Increases in dose at close range due to the presence of the shield are observed. Such enhancement is in agreement with Williamson's Monte Carlo calculations

Additional details

Publishing Information

Journal Title
Med. Phys.
Journal Volume
15
Journal Issue
4
Series
Med. Phys.
Journal Page Range
627-628
ISSN
0094-2405
CODEN
MPHYA