Published December 2002 | Version v1
Journal article

Accuracy and initial clinical experience with measurement software (advanced vessel analysis) in three-dimensional imaging

  • 1. Kurume Univ., Fukuoka (Japan). School of Medicine
  • 2. Kurume Univ., Fukuoka (Japan). Hospital

Description

Recently, the clinical benefits of three dimensional (3D) imaging, such as 3D-CTA and 3D-DSA, in cerebro-vascular disease have been widely recognized. Software for quantitative analysis of vascular structure in 3D imaging (advanced vessel analysis: AVA) has been developed. We evaluated AVA with both phantom studies and a few clinical cases. In spiral and curvy aluminum tube phantom studies, the accuracy of diameter measurements was good in 3D images produced from data set generated by multi-detector row CT or rotational angiography. The measurement error was less than 0.03 mm on aluminum tube phantoms that were 3 mm and 5 mm in diameter. In the clinical studies, the differences of carotid artery diameter measurements between 2D-DSA and 3D-DSA was less than 0.3 mm in. The measurement of length, diameter and angle by AVA should provide useful information for planning surgical and endovascular treatments of cerebro-vascular disease. (author)

Additional details

Publishing Information

Journal Title
CI Kenkyu
Journal Volume
24
Journal Issue
4
Journal Page Range
p. 215-220
ISSN
0918-7073