Published 1990 | Version v1
Book

The long-term stability and reliability of photomultipliers used in Anger scintillation cameras

Creators

  • 1. Siemens Gammasonics, Inc., Hoffman Estates, IL (USA)

Description

A test instrument has been designed to measure the stability of PMT's (photomultiplier tubes) used in Anger scintillation cameras. The instrument is comprised of a number of glass capsules containing beta-emitting tritium with a blue phosphor to provide very stable light sources along with a data logger to monitor anode current. Three hundred sixty PMT's are simultaneously tested for anode luminous sensitivity, dark current, long-term drift rate, erratic gain instabilities and rms noise. This paper discusses results in the use of the instrument to test over 300 000 PMT's and the impact upon Anger camera reliability is assessed

Additional details

Publishing Information

Publisher
IEEE Service Center.
Imprint Place
Piscataway, NJ (USA)
Imprint Title
IEEE nuclear science symposium conference record emdash 1990
Imprint Pagination
1636 p.
Journal Page Range
p. 163-168.

Conference

Title
1990 Institute of Electrical and Electronics Engineers (IEEE) nuclear science symposium.
Dates
22-27 Oct 1990.
Place
Arlington, VA (USA).

Optional Information

Secondary number(s)
CONF-9010220--.