NPL contribution to an international comparison of 24.5 ke V neutron fluence measurements

Description

This report contains details of the experimental procedure and the analysis techniques used by NPL during an international comparison of neutron fluence production and measurement in the energy region around 24 keV. The exercise involved the calibration of three spheres from a Bonner sphere set in terms of their fluence response to 27.4 and 24.6 keV neutrons produced using the 45Sc(p,n)45Ti reaction. (author)

Availability note (English)

Available from British Library Document Supply Centre- DSC:6029.28071(46)

Additional details

Publishing Information

Imprint Pagination
27 p.
Journal Issue
no.CIRM46
Series
NPL report
Report number
NPL-R-CIRM--46