Combined scintillation and track technique to search for ββ-decay
Description
A variant of the scintillation spectrometer with wire chambers for ββ-decay search is considered. The time-of-flight technique for background suppression is suggested. The (-3 m2) source placed in the centre of the facility comprises three planes. Four wire chambers measure particle tracks in the centre of the source. The whole volume is filled with helium. Scintillation counters are produced of long plates. Four plates form a counter with two fast photomultipliers on both edges. Electron time-of-flight between the scintillators makes up 2-5 ns. If the mass of the (150Nd) source is equal to 75 g, then the half-decay period may be measured T1/2=1022 years
Availability note (English)
MF available from INIS under the Report Number.Files
20019528.pdf
Files
(121.8 kB)
| Name | Size | Download all |
|---|---|---|
|
md5:0cb275940c8f45627f3c4dca980057a3
|
121.8 kB | Preview Download |
Additional details
Publishing Information
- Imprint Pagination
- 4 p.
- Report number
- JINR-E--13-87-309
INIS
- Country of Publication
- USSR
- Country of Input or Organization
- USSR
- INIS RN
- 20019528
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- BACKGROUND RADIATION; DOUBLE BETA DECAY; EFFICIENCY; HALF-LIFE; MULTIWIRE IONIZATION CHAMBERS; NEODYMIUM 150; SOLID SCINTILLATION DETECTORS; TIME RESOLUTION; TIME-OF-FLIGHT SPECTROMETERS
- Descriptors DEC
- BETA DECAY; BETA-MINUS DECAY; DECAY; EVEN-EVEN NUCLEI; INTERMEDIATE MASS NUCLEI; IONIZATION CHAMBERS; ISOTOPES; MEASURING INSTRUMENTS; NEODYMIUM ISOTOPES; NUCLEAR DECAY; NUCLEI; RADIATION DETECTORS; RADIATIONS; RARE EARTH NUCLEI; RESOLUTION; SCINTILLATION COUNTERS; SPECTROMETERS; STABLE ISOTOPES; TIMING PROPERTIES
Optional Information
- Notes
- 6 refs.; 1 fig.; submitted to the journal NIM.