The development of a Compton lung densitometer
Description
A field instrument is being developed for the non-invasive determination of absolute lung density using unique Compton backscattering techniques. A system consisting of a monoenergetic gamma-ray beam and a shielded high resolution high-purity-germanium (HPGe) detector in a close-coupled geometry is designed to minimize errors due to multiple scattering and uncontrollable attenuation in the chestwall. Results of studies on system performance with phantoms, the optimization of detectors, and the fabrication of a practical gamma-ray source are presented. 3 refs., 6 figs., 2 tabs
Availability note (English)
MF available from INIS under the Report Number; Available from NTIS, PC A03/MF A01 - OSTI; 1 as DE89006654.
Files
Additional details
Publishing Information
- Imprint Pagination
- 18 p.
- Report number
- LBL--25170
Conference
- Title
- IEEE nuclear science symposium.
- Dates
- 9-11 Nov 1988.
- Place
- Orlando, FL (USA).
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 20037226
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- COMPTON DIODE DETECTORS; DENSITY; DETECTION; DIAGNOSIS; EDEMA; GAMMA RADIATION; GOATS; LUNGS; MONITORING; OPTIMIZATION; PHANTOMS; PORTABLE EQUIPMENT; RESPIRATORY SYSTEM DISEASES; SHEEP
- Descriptors DEC
- ANIMALS; BODY; DISEASES; DOMESTIC ANIMALS; ELECTROMAGNETIC RADIATION; EQUIPMENT; IONIZING RADIATIONS; MAMMALS; MEASURING INSTRUMENTS; MOCKUP; ORGANS; PATHOLOGICAL CHANGES; PHYSICAL PROPERTIES; RADIATION DETECTORS; RADIATIONS; RESPIRATORY SYSTEM; RUMINANTS; STRUCTURAL MODELS; SYMPTOMS; VERTEBRATES