Published December 2010 | Version v1
Journal article

Energy dependence of three passive solid detectors on low energy X-ray

  • 1. National Institute for radiological Protection, Chinese Center for Disease Control and Prevention, Beijing (China)

Description

Objective: To assess the feasibility of ESD measurements in the mammography by the comparative study of energy dependence of the three passive solid detectors on low energy X-ray. Methods: Two thermoluminescent TL) detectors (LiF (Mg, Ti) and LiF (Mg, Cu, P) and one optically stimulated luminescence (OSL) detector (Al2O3 : C) were irradiated for the same air kerma by ten kinds of beam quality between 25-40 kV in a Mo/Mo low energy X-ray generator,respectively. Results: The energy dependences of TL and OSL were 25% and 11% between 25-40 kV, respectively. Conclusions: considerable potential for OSL detector can be used in routine quality control and ESD measurements in the mammography. (authors)

Additional details

Publishing Information

Journal Title
Chinese Journal of Radiological Medicine and Protection
Journal Volume
30
Journal Issue
6
Journal Page Range
p. 735-736
ISSN
0254-5098

Optional Information

Notes
1 figs., 5 refs.