Published September 7, 2010 | Version v1
Journal article

Evaluation of image reconstruction for mouse brain imaging with synthetic collimation from highly multiplexed SiliSPECT projections

  • 1. Department of Radiology and Radiological Sciences, Vanderbilt University Institute of Imaging Science, Vanderbilt University, Nashville, TN (United States)
  • 2. Center for Gamma-Ray Imaging, Department of Radiology, University of Arizona, Tucson, AZ (United States)
  • 3. Department of Radiology, University of Pennsylvania, Philadelphia, PA (United States)

Description

We have performed a theoretical study to explore the potential and limitations of synthetic collimation for SPECT imaging with stacked-detector acquisition (dual magnification). This study will be used to optimize SiliSPECT, a small-animal SPECT for imaging small volumes such as a mouse brain at high sensitivity and resolution. The synthetic collimation enables image reconstruction with a limited number of camera views and in the presence of significant multiplexing. We also developed a new formulation to quantify the multiplexed object sensitivity and investigated how this changes for different acquisition parameters such as number of pinholes and combinations of front and back detector distances for imaging objects as small as the mouse brain. In our theoretical studies, we were not only able to demonstrate better reconstruction results by incorporating two detector magnifications in comparison to either one alone, but also observed an improved image reconstruction by optimizing the detector-collimator distances to change the multiplexing ratio between the front and back detectors.

Availability note (English)

Available from http://dx.doi.org/10.1088/0031-9155/55/17/017

Additional details

Identifiers

DOI
10.1088/0031-9155/55/17/017;
PII
S0031-9155(10)48338-4;

Publishing Information

Journal Title
Physics in Medicine and Biology
Journal Volume
55
Journal Issue
17
Journal Page Range
p. 5151-5168
ISSN
0031-9155
CODEN
PHMBA7

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
42025211
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
BRAIN; COLLIMATORS; IMAGE PROCESSING; MICE; OPTIMIZATION; SENSITIVITY; SINGLE PHOTON EMISSION COMPUTED TOMOGRAPHY
Descriptors DEC
ANIMALS; BODY; CENTRAL NERVOUS SYSTEM; COMPUTERIZED TOMOGRAPHY; DIAGNOSTIC TECHNIQUES; EMISSION COMPUTED TOMOGRAPHY; MAMMALS; NERVOUS SYSTEM; ORGANS; PROCESSING; RODENTS; TOMOGRAPHY; VERTEBRATES