Published 1987 | Version v1
Journal article

Theory and capabilities of a 4πβ-γ coincidence system with cumulative dead times

  • 1. CEA Centre d'Etudes Nucleaires de Saclay, 91 - Gif-sur-Yvette (France). Lab. de Metrologie des Rayonnements Ionisants

Description

A 4πβ-γ coincidence system working with cumulative dead times, and associated electronics, is described. The corresponding theoretical formulae are presented. For testing both electronics and formulae, high-activity sources of 60Co (up to 1.2 MBq) have been measured successfully. Their estimated uncertainties at one standard deviation were less than 0.25 %. (author)

Additional details

Publishing Information

Journal Title
Appl. Radiat. Isot.
Journal Volume
38
Journal Issue
10
Series
Appl. Radiat. Isot.
Journal Page Range
799-804
CODEN
ARISE

Conference

Title
ICRM seminar on techniques in radionuclide metrology.
Dates
15-16 Jun 1987.
Place
Rome (Italy).