Published December 19, 2012 | Version v1
Journal article

Routine 18F− production with 180 μA to 200 μA target beam current on a GE PETtrace 800 cyclotron

  • 1. Department of PET and Nuclear Medicine, Royal Prince Alfred Hospital, Sydney, NSW (Australia)
  • 2. GEMS PET Systems AB, Uppsala (Sweden)

Description

Through upgrades and enhancements, which are now standard on current PETtrace 800 cyclotrons, a GE PETtrace cyclotron installed in 2002 has demonstrated reliable routine [18F]FDG production at total target beam currents of 180 μA without detriment to saturation and [18F]FDG yields. Routine production at 200 μA has been achieved and its evaluation is continuing. Self-shielded target using W/Cu alloy for the target body afforded a reduction in dose rate from the Havar foils by a factor of ∼ 8-10, reducing dose from the targets and need for removing targets during maintenance. The main activation product in the shield is 187W (T1/2 24 h). The 60Co ((T1/2 5.3 y) activation is about 250 times less at 24 h post EOB and is not considered a major issue despite its long half-life.

Additional details

Identifiers

Publishing Information

Journal Title
AIP Conference Proceedings
Journal Volume
1509
Journal Issue
1
Journal Page Range
p. 66-70
ISSN
0094-243X
CODEN
APCPCS

Conference

Title
14. international workshop on targetry and target chemistry
Dates
26-29 Aug 2012
Place
Playa del Carmen (Mexico)

Optional Information

Notes
(c) 2012 American Institute of Physics