A novel procedure for determining the optimal MLC configuration parameters in treatment planning systems based on measurements with a Farmer chamber

  • 1. Department of Radiation Oncology, Hospital Clínic de Barcelona, 08036 Barcelona (Spain)
  • 2. Department of Medical Physics, Hospital Universitari Sant Joan de Reus, IISPV, 43204 Tarragona (Spain)
  • 3. Iridium Kankernetwerk, Antwerp (Belgium)

Description

Modelling of the multi-leaf collimator (MLC) in treatment planning systems (TPS) is crucial for the dose calculation accuracy of intensity-modulated radiation therapy plans. However, no standardised methodology for their configuration exists to date. In this study we present a method that separates the effect of each dosimetric characteristic of the MLC, offering comprehensive equations for the determination of the configuration parameters used in the TPS model. The main advantage of the method is that it only requires prior knowledge of the nominal leaf width and is based on doses measured with a Farmer chamber, which is a very well established and robust methodology. Another significant advantage is the required time, since measuring the tests takes only about 30 minutes per energy. Firstly, we provide a theoretical general formalism in terms of the primary fluence constructed from the transmission map obtained from an MLC model for synchronous and asynchronous sweeping beams. Secondly, we apply the formalism to the RayStation TPS as a proof of concept and we derive analytical expressions that allow the determination of the configuration parameters (leaf tip width, tongue-and-groove width, x-position offset and MLC transmission) and describe how they intertwine. Finally, we apply the method to Varian's Millennium120 and HD120 MLCs in a TrueBeam linear accelerator for different energies and determine the optimal configuration parameters. The proposed procedure is much faster and streamlined than the typical trial-and-error methods and increases the accuracy of dose calculation in clinical plans. Additionally, the procedure can be useful for standardising the MLC configuration process and it exposes the limitations of the implemented MLC model, providing guidance for further improvement of these models in TPSs. (paper)

Availability note (English)

Available from http://dx.doi.org/10.1088/1361-6560/ab8cd5

Additional details

Identifiers

Publishing Information

Journal Title
Physics in Medicine and Biology
Journal Volume
65
Journal Issue
15
Journal Page Range
[13 p.]
ISSN
0031-9155
CODEN
PHMBA7

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
52074185
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
ACCURACY; COLLIMATORS; LINEAR ACCELERATORS; PLANNING; RADIOTHERAPY; SIMULATION; THERAPEUTIC DOSES
Descriptors DEC
ACCELERATORS; DOSES; MEDICINE; NUCLEAR MEDICINE; RADIOLOGY; THERAPY