Dose reduction in maxillofacial imaging using low dose Cone Beam CT

  • 1. Department of Oral Diagnosis and Oral Radiology, School of Dentistry, University of Athens (Greece)
  • 2. Licensing and Inspection Division, Greek Atomic Energy Commission (Greece)

Description

Objectives: (a) To measure the absorbed dose at certain anatomical sites of a RANDO phantom and to estimate the effective dose in radiographic imaging of the jaws using low dose Cone Beam computed tomography (CBCT) and (b) to compare the absorbed and the effective doses between thyroid and cervical spine shielding and non-shielding techniques. Study design: Thermoluminescent dosimeters (TLD-100) were placed at 14 sites in a RANDO phantom, using a Cone Beam CT device (Newtom, Model QR-DVT 9000, Verona, Italy). Dosimetry was carried out applying two techniques: in the first, there was no shielding device used while in the second one, a shielding device (EUREKA!, TRIX) was applied for protection of the thyroid gland and the cervical spine. Effective dose was estimated according to ICRP6 report (E ICRP). An additional estimation of the effective dose was accomplished including the doses of the salivary glands (E SAL). A Wilcoxon Signed Ranks Test was used for statistical analysis. Results: In the non-shielding technique the absorbed doses ranged from 0.16 to 1.67 mGy, while 0.32 and 1.28 mGy were the doses to the thyroid and the cervical spine, respectively. The effective dose, E ICRP, was 0.035 mSv and the E SAL was 0.064 mSv. In the shielding technique, the absorbed doses ranged from 0.09 to 1.64 mGy, while 0.18 and 0.95 mGy were the respective values for the thyroid and the cervical spine. The effective dose, E ICRP, was 0.023 mSv and E SAL was 0.052 mSv. Conclusions: The use of CBCT for maxillofacial imaging results in a reduced absorbed and effective dose. The use of lead shielding leads to a further reduction of the absorbed doses of thyroid and cervical spine, as well as the effective dose

Additional details

Identifiers

DOI
10.1016/j.ejrad.2005.05.011;
PII
S0720-048X(05)00198-1;

Publishing Information

Journal Title
European Journal of Radiology
Journal Volume
56
Journal Issue
3
Journal Page Range
p. 413-417
ISSN
0720-048X
CODEN
EJRADR

INIS

Optional Information

Copyright
Copyright (c) 2005 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.