Published December 1990 | Version v1
Journal article

An amplifier for use with large-area photodiodes

  • 1. Fusion Energy Division, Oak Ridge National Laboratory, Oak Ridge, TN (USA)

Description

We describe a high-gain (∼107) amplifier for use with large-area silicon photodiodes for monitoring the intensity of plasma light from fusion experiments. To achieve the necessary gain without saturating the output signal, the amplifier design incorporates a bootstrap technique and capacitive coupling between amplifier stages. This design eliminates voltage offset at the amplifier output due to detector leakage, while retaining desirable low-frequency response characteristics

Additional details

Publishing Information

Journal Title
Review of Scientific Instruments
Journal Volume
61
Journal Issue
12
Series
Rev. Sci. Instrum.
Journal Page Range
3909-3910
ISSN
0034-6748
CODEN
RSINA