Published March 1998
| Version v1
Report
Development of twin Ge detector for high energy photon measurement and its performance
Creators
- 1. Power Reactor and Nuclear Fuel Development Corp., Tokai, Ibaraki (Japan). Tokai Works
Description
Prototype twin HPGe detector composed of two large HPGe crystals was developed to obtain better detection efficiency (ε) and P/T ratio, which was required for high energy photon spectroscopy. In this work, the performances of the twin HPGe detector were evaluated by computer simulation employing EGS4 code. (author)
Additional details
Publishing Information
- Imprint Title
- Proceedings of the 1997 symposium on nuclear data
- Imprint Pagination
- 372 p.
- Journal Page Range
- p. 307-311
- Report number
- JAERI-Conf--98-003
Conference
- Title
- 1997 symposium on nuclear data
- Dates
- 27-28 Nov 1997
- Place
- Tokai, Ibaraki (Japan)
INIS
- Country of Publication
- Japan
- Country of Input or Organization
- Japan
- INIS RN
- 29053526
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- E CODES; EFFICIENCY; HIGH-PURITY GE DETECTORS; MEV RANGE 10-100; MONTE CARLO METHOD; PHOTOELECTRIC EFFECT; PROTON SPECTROMETERS
- Descriptors DEC
- CALCULATION METHODS; COMPUTER CODES; ENERGY RANGE; GE SEMICONDUCTOR DETECTORS; MEASURING INSTRUMENTS; MEV RANGE; RADIATION DETECTORS; SEMICONDUCTOR DETECTORS; SPECTROMETERS
Optional Information
- Secondary number(s)
- INDC(JPN)--180/U