Published October 2019 | Version v1
Journal article

Development and preliminary results of Xtrim-PET, a modular cost-effective preclinical scanner

  • 1. Gordon Center for Medical Imaging, Department of Radiology, Massachusetts General Hospital and Harvard Medical School, Boston, MA 02114 (United States)
  • 2. Department of Radiology, University of Massachusetts Medical School, Worcester, MA (United States)
  • 3. BlueIce Corporation, Brussels (Belgium)
  • 4. Research Center for Molecular and Cellular Imaging, Tehran University of Medical Sciences, Tehran (Iran, Islamic Republic of)
  • 5. Research Center for Science and Technology in Medicine, Tehran University of Medical Sciences, Tehran (Iran, Islamic Republic of)
  • 6. Department of Medical Physics and Biomedical Engineering, Tehran University of Medical Sciences, Tehran (Iran, Islamic Republic of)

Description

We report on development of Xtrim-PET, a prototype low-cost preclinical PET system using high-density analog silicone photomultiplier (SiPM) arrays and pixelated scintillator LYSO:Ce crystals. Channel reduction method is employed to reduce 144 SiPM signals per detector block to 16 energy and 9 fast signals in the detector head. Using these signals, a Digital Front-End (DFE) board calculates energy, position, and time information of incident gamma photon and transmits data packets to a downstream Digital Coincidence Processing Unit (DCPU) in which the coincidence events are extracted in a fully pipelined digital processing fashion for maximum count throughput. The system consists of 10 detector blocks (detector head and DFE board), a DCPU board, and an acquisition computer. Prompt and delayed coincidence events are stored in list mode in the acquisition computer where first hand corrections are applied and 3D sinogram is generated in offline mode. Two-dimensional sinograms are then generated using single-slice rebinning algorithm and iterative image reconstruction methods are employed to generate the 3D image. The developed scanner works in room temperature without additional cooling system and has 180 mm ring diameter. Tomographic image resolution of 1.6, 1.7, and 1.8 mm FWHM was calculated along radial, tangential, and axial directions, respectively. Preliminary investigations were performed using NEMA image quality phantom in addition to rat cardiac and bone scans for preclinical evaluation of the scanner.

Additional details

Identifiers

DOI
10.1016/j.nima.2019.06.009;
PII
S0168900219308289;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
940
Journal Page Range
p. 288-295
ISSN
0168-9002
CODEN
NIMAER

Optional Information

Copyright
Copyright (c) 2019 Elsevier B.V. All rights reserved.