Pencil Beam Algorithms Are Unsuitable for Proton Dose Calculations in Lung
Description
Purpose: To compare analytic and Monte Carlo–based algorithms for proton dose calculations in the lung, benchmarked against anthropomorphic lung phantom measurements. Methods and Materials: A heterogeneous anthropomorphic moving lung phantom has been irradiated at numerous proton therapy centers. At 5 centers the treatment plan could be calculated with both an analytic and Monte Carlo algorithm. The doses calculated in the treatment plans were compared with the doses delivered to the phantoms, which were measured using thermoluminescent dosimeters and film. Point doses were compared, as were planar doses using a gamma analysis. Results: The analytic algorithms overestimated the dose to the center of the target by an average of 7.2%, whereas the Monte Carlo algorithms were within 1.6% of the physical measurements on average. In some regions of the target volume, the analytic algorithm calculations differed from the measurement by up to 31% in the internal gross target volume (iGTV) (46% in the planning target volume), over-predicting the dose. All comparisons showed a region of at least 15% dose discrepancy within the iGTV between the analytic calculation and the measured dose. The Monte Carlo algorithm recalculations showed dramatically improved agreement with the measured doses, showing mean agreement within 4% for all cases and a maximum difference of 12% within the iGTV. Conclusions: Analytic algorithms often do a poor job predicting proton dose in lung tumors, over-predicting the dose to the target by up to 46%, and should not be used unless extensive validation counters the consistent results of the present study. Monte Carlo algorithms showed dramatically improved agreement with physical measurements and should be implemented to better reflect actual delivered dose distributions.
Availability note (English)
Available from http://dx.doi.org/10.1016/j.ijrobp.2017.06.003Additional details
Identifiers
- DOI
- 10.1016/j.ijrobp.2017.06.003;
- PII
- S0360-3016(17)31012-X;
Publishing Information
- Journal Title
- International Journal of Radiation Oncology, Biology and Physics
- Journal Volume
- 99
- Journal Issue
- 3
- Journal Page Range
- p. 750-756
- ISSN
- 0360-3016
- CODEN
- IOBPD3
INIS
- Country of Publication
- United States
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 49073845
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- FILM DOSIMETRY; LUNGS; MONTE CARLO METHOD; PHANTOMS; PROTON BEAMS; RADIATION DOSE DISTRIBUTIONS; RADIATION DOSES; THERMOLUMINESCENT DOSEMETERS
- Descriptors DEC
- BEAMS; BODY; CALCULATION METHODS; DOSEMETERS; DOSES; DOSIMETRY; LUMINESCENT DOSEMETERS; MEASURING INSTRUMENTS; MOCKUP; NUCLEON BEAMS; ORGANS; PARTICLE BEAMS; RESPIRATORY SYSTEM; STRUCTURAL MODELS
Optional Information
- Copyright
- Copyright (c) 2017 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.