Published November 2017 | Version v1
Journal article

Statistical process control analysis for patient quality assurance of intensity modulated radiation therapy

  • 1. College of Medicine Ewha Womans University, Department of radiation Oncology (Korea, Republic of)
  • 2. Kosin University College of Medicine, Department of Radiation Oncology (Korea, Republic of)
  • 3. College of Medicine, Korea University, Department of Radiation Oncology (Korea, Republic of)
  • 4. Inha University Hospital, Department of Radiation Oncology (Korea, Republic of)
  • 5. Cheil General Hospital & Women's Heathcare Center, Department of Radiation Oncology (Korea, Republic of)
  • 6. College of Medicine, Yonsei University, Department of Radiation Oncology (Korea, Republic of)

Description

This study applied statistical process control to set and verify the quality assurances (QA) tolerance standard for our hospital's characteristics with the criteria standards that are applied to all the treatment sites with this analysis. Gamma test factor of delivery quality assurances (DQA) was based on 3%/3 mm. Head and neck, breast, prostate cases of intensity modulated radiation therapy (IMRT) or volumetric arc radiation therapy (VMAT) were selected for the analysis of the QA treatment sites. The numbers of data used in the analysis were 73 and 68 for head and neck patients. Prostate and breast were 49 and 152 by MapCHECK and ArcCHECK respectively. Cp value of head and neck and prostate QA were above 1.0, Cpml is 1.53 and 1.71 respectively, which is close to the target value of 100%. Cpml value of breast (IMRT) was 1.67, data values are close to the target value of 95%. But value of was 0.90, which means that the data values are widely distributed. Cp and Cpml of breast VMAT QA were respectively 1.07 and 2.10. This suggests that the VMAT QA has better process capability than the IMRT QA. Consequently, we should pay more attention to planning and QA before treatment for breast Radiotherapy.

Additional details

Identifiers

Publishing Information

Journal Title
Journal of the Korean Physical Society
Journal Volume
71
Journal Issue
10
Journal Page Range
p. 717-721
ISSN
0374-4884
CODEN
KPSJAS

INIS

Country of Publication
Korea, Republic of
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
54093404
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
MAMMARY GLANDS; PATIENTS; PROCESS CONTROL; PROSTATE; QUALITY ASSURANCE; RADIOTHERAPY
Descriptors DEC
BODY; CONTROL; GLANDS; MALE GENITALS; MANAGEMENT; MEDICINE; NUCLEAR MEDICINE; ORGANS; QUALITY MANAGEMENT; RADIOLOGY; THERAPY

Optional Information

Copyright
Copyright (c) 2017 The Korean Physical Society