Description
Compton suppression spectrometer for the measurement of low background γ ray has adapted the anti-coincidence counting system of parallel sampling model and an effective material shielding, also automatic data collecting and processing. In the well-type anti-coincidence model, the detecting sensitivity was 2.02 pCi for 137Cs standard thin-film source which was placed above 3.5 cm from NaI(Tl) main detector. Integral background was 115 cpm in the energy range 50 ∼ 2000 keV. Compared to the model without anti-coincidence, it was reduced by 61.5%
Additional details
Publishing Information
- Journal Title
- Nuclear Electronics and Detection Technology
- Journal Volume
- 15
- Journal Issue
- 3
- Journal Page Range
- p. 172-175, 197.
- ISSN
- 0258-0934
- CODEN
- HDYUEC
INIS
- Country of Publication
- China
- Country of Input or Organization
- China
- INIS RN
- 26067276
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- ANTICOINCIDENCE; AUTOMATION; BACKGROUND NOISE; COMPARATIVE EVALUATIONS; COMPTON EFFECT; DATA ACQUISITION; DATA PROCESSING; GAMMA SPECTROMETERS; PERFORMANCE; SENSITIVITY; SHIELDING; STABILITY
- Descriptors DEC
- BASIC INTERACTIONS; ELASTIC SCATTERING; ELECTROMAGNETIC INTERACTIONS; EVALUATION; INTERACTIONS; MEASURING INSTRUMENTS; NOISE; SCATTERING; SPECTROMETERS