A tandem system performance for quality control in radiodiagnostic range
Creators
- 1. Instituto de Pesquisas Energeticas e Nucleares (IPEN/CNEN-SP), Sao Paulo, SP (Brazil)
Description
A tandem ionization chamber system has been developed to be applied for conventional diagnostic radiology energy range dosimetry. This system consists of two ionization chambers with different collecting electrode materials: aluminum and graphite. Both chambers have a sensitive volume of 6 cm3. Characteristics as saturation, ion collection efficiency, linearity of chamber response versus air kerma rate and energy dependence were determined. The results of energy dependence allowed the determination of the tandem curve. It can be used in quality control programs of diagnostic radiology area. All measurements were carried out at the Calibration Laboratory of IPEN. (author)
Additional details
Additional titles
- Original title (Portuguese)
- Desempenho de um sistema tandem para controle de qualidade em radiodiagnostico
Identifiers
Publishing Information
- Journal Title
- Revista Brasileira de Fisica Medica (Online)
- Journal Volume
- 4
- Journal Issue
- 3
- Journal Page Range
- p. 27-30
- ISSN
- 1984-9001
INIS
- Country of Publication
- Brazil
- Country of Input or Organization
- Brazil
- INIS RN
- 42073029
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- AVALANCHE QUENCHING; DIAGNOSIS; DIAGNOSTIC TECHNIQUES; DOSEMETERS; DOSIMETRY; ELECTRON-CAPTURE DETECTORS; IONIZATION CHAMBERS; MEDICAL EXAMINATIONS; RADIATION DETECTION; RADIATION DOSES; RADIATIONS; RADIOLOGY
- Descriptors DEC
- DETECTION; DOSES; MEASURING INSTRUMENTS; MEDICAL SURVEILLANCE; MEDICINE; NUCLEAR MEDICINE; RADIATION DETECTORS; RADIOMETRIC GAGES
- Proposed descriptors and Free-text terms
- IONIZATION CHAMBER; RADIODIAGNOSTIC; TANDEM SYSTEM; INSTRUMENTATION
Optional Information
- Notes
- The journal lists Vol. 4(3) 2010 in the table of contents, while the PDF is Vol. 4(3) 2011.