A tandem system performance for quality control in radiodiagnostic range

  • 1. Instituto de Pesquisas Energeticas e Nucleares (IPEN/CNEN-SP), Sao Paulo, SP (Brazil)

Description

A tandem ionization chamber system has been developed to be applied for conventional diagnostic radiology energy range dosimetry. This system consists of two ionization chambers with different collecting electrode materials: aluminum and graphite. Both chambers have a sensitive volume of 6 cm3. Characteristics as saturation, ion collection efficiency, linearity of chamber response versus air kerma rate and energy dependence were determined. The results of energy dependence allowed the determination of the tandem curve. It can be used in quality control programs of diagnostic radiology area. All measurements were carried out at the Calibration Laboratory of IPEN. (author)

Additional details

Additional titles

Original title (Portuguese)
Desempenho de um sistema tandem para controle de qualidade em radiodiagnostico

Publishing Information

Journal Title
Revista Brasileira de Fisica Medica (Online)
Journal Volume
4
Journal Issue
3
Journal Page Range
p. 27-30
ISSN
1984-9001

Optional Information

Notes
The journal lists Vol. 4(3) 2010 in the table of contents, while the PDF is Vol. 4(3) 2011.