Automatic image segmentation based on synthetic tissue model for delineating organs at risk in spinal metastasis treatment planning
Creators
- 1. Department of Radiation Oncology, Universitätsklinikum Schleswig-Holstein Campus Kiel (Germany)
- 2. R&D Anatomical Mapping, Brainlab AG, Munich (Germany)
Description
One of the main goals in software solutions for treatment planning is to automatize delineation of organs at risk (OARs). In this pilot feasibility study a clinical validation was made of computed tomography (CT)-based extracranial auto-segmentation (AS) using the Brainlab Anatomical Mapping tool (AM). The delineation of nine extracranial OARs (lungs, kidneys, trachea, heart, liver, spinal cord, esophagus) from clinical datasets of 24 treated patients was retrospectively evaluated. Manual delineation of OARs was conducted in clinical routine and compared with AS datasets using AM. The Dice similarity coefficient (DSC) and maximum Hausdorff distance (HD) were used as statistical and geometrical measurements, respectively. Additionally, all AS structures were validated using a subjective qualitative scoring system. All patient datasets investigated were successfully processed with the evaluated AS software. For the left lung (0.97 ± 0.03), right lung (0.97 ± 0.05), left kidney (0.91 ± 0.07), and trachea (0.93 ± 0.04), the DSC was high with low variability. The DSC scores of other organs (right kidney, heart, liver, spinal cord), except the esophagus, ranged between 0.7 and 0.9. The calculated HD values yielded comparable results. Qualitative assessment showed a general acceptance in more than 85% of AS OARs - except for the esophagus. The Brainlab AM software is ready for clinical use in most of the OARs evaluated in the thoracic and abdominal region. The software generates highly conformal structure sets compared to manual contouring. The current study design needs revision for further research.
Availability note (English)
Available from: http://dx.doi.org/10.1007/s00066-019-01463-4Additional details
Identifiers
Publishing Information
- Journal Title
- Strahlentherapie und Onkologie
- Journal Volume
- 195
- Journal Issue
- 12
- Journal Page Range
- p. 1094-1103
- ISSN
- 0179-7158
- CODEN
- STONE4
INIS
- Country of Publication
- Germany
- Country of Input or Organization
- Germany
- INIS RN
- 51014733
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- COMPARATIVE EVALUATIONS; COMPUTER CODES; COMPUTERIZED SIMULATION; COMPUTERIZED TOMOGRAPHY; ESOPHAGUS; EXTERNAL BEAM RADIATION THERAPY; FEASIBILITY STUDIES; KIDNEYS; LIVER; LUNGS; METASTASES; PLANNING; SPATIAL DOSE DISTRIBUTIONS; SPINAL CORD; TRACHEA; VALIDATION
- Descriptors DEC
- BODY; CENTRAL NERVOUS SYSTEM; DIAGNOSTIC TECHNIQUES; DIGESTIVE SYSTEM; EVALUATION; GLANDS; MEDICINE; NERVOUS SYSTEM; NUCLEAR MEDICINE; ORGANS; RADIATION DOSE DISTRIBUTIONS; RADIOLOGY; RADIOTHERAPY; RESPIRATORY SYSTEM; SIMULATION; TESTING; THERAPY; TOMOGRAPHY