Published March 24, 2008 | Version v1
Journal article

A fluorescent chemosensor for calcium with excellent storage stability in water

  • 1. OPTI Medical Systems, Inc., 235 Hembree Park Drive, Roswell, GA 30076 (United States)

Description

This article describes the design, synthesis and characterization of an optical sensor suitable for practical measurement of ionized calcium in serum and whole blood samples. The key to the development of this sensor is the identification of a chemically very stable, nitrogen-containing, calcium selective ionophore, coupled with a fluorophore having the correct spectral and electron accepting properties. The slope of the sensor is about 34%/mM in the typical clinically significant range of 0.32-2.2 mM. This sensor has been implemented into the disposable cartridge, used for commercially available OPTI CCA analyzer with precision better than ±0.02 mM (1 S.D.). The sensor displays excellent stability against hydrolysis and oxidation, leading to less than 0.02 mM measurement error after 9 months of wet storage at room temperature, up to 30 deg. C

Availability note (English)

Available from http://dx.doi.org/10.1016/j.aca.2008.01.059

Additional details

Identifiers

DOI
10.1016/j.aca.2008.01.059;
PII
S0003-2670(08)00238-9;

Publishing Information

Journal Title
Analytica Chimica Acta
Journal Volume
611
Journal Issue
2
Journal Page Range
p. 197-204
ISSN
0003-2670
CODEN
ACACAM

Optional Information

Copyright
Copyright (c) 2008 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.