Published February 1, 2007 | Version v1
Journal article

Evaluation of tomographic reconstruction methods for small animal microCT and microPET/CT

  • 1. CREATIS, UMR CNRS 5515, 69621 Villeurbanne (France) and CREATIS, U630 INSERM, 69621 Villeurbanne (France) and CREATIS, INSA-Lyon, 69621 Villeurbanne (France) and CREATIS, Universite Claude Bernard-Lyon 1, 69677 Bron (France)
  • 2. CREATIS, Universite Claude Bernard-Lyon 1, 69677 Bron (France)
  • 3. CREATIS, INSA-Lyon, 69621 Villeurbanne (France)
  • 4. CREATIS, U630 INSERM, 69621 Villeurbanne (France)
  • 5. CREATIS, UMR CNRS 5515, 69621 Villeurbanne (France)
  • 6. CPPM, CNRS, IN2P3 UMR6550, Universite de la Mediterranee-Aix-Marseille II, 13288 Marseille (France)

Description

Biological research on small animals is constantly demanding for imaging devices with higher performances like combined scanners enabling multiple modality data acquisition. In this context, we consider the concept of using hybrid X-ray pixel detectors (XPAD) with the clearPET to build a prototype of microPET/CT. The system will permit simultaneous acquisition of the anatomical and functional modalities thanks to the placement of the XPAD in front of the gamma detectors. It is foreseen to place the X-ray source outside the PET ring, thus involving a non-conventional off-centered circular geometry for the CT data acquisition. In a first step, a series of acquisitions were performed with a microCT, built with the XPAD2 detector. Our work consists first, in the investigation of several CT reconstruction methods for these acquisitions. We hereby compare the reconstruction results obtained with two variants of the Feldkamp and two analytical methods developed to compensate for some artifacts. Second, the reconstruction issue in the off-centered geometry is addressed by testing the same algorithms on numerical simulations. It appears that, if the next generation of XPAD detectors using smaller pixels is needed to obtain high resolution images, the reconstruction method employed plays an important role and can efficiently improve the quality of the results

Additional details

Identifiers

DOI
10.1016/j.nima.2006.10.082;
PII
S0168-9002(06)01879-1;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
571
Journal Issue
1-2
Journal Page Range
p. 278-281
ISSN
0168-9002
CODEN
NIMAER

Conference

Title
1. international conference on molecular imaging technology
Acronym
EuroMedIm 2006
Dates
9-12 May 2006
Place
Marseille (France)

INIS

Country of Publication
Netherlands
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
38092108
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
ALGORITHMS; ANIMALS; BEAMS; COMPARATIVE EVALUATIONS; COMPUTERIZED SIMULATION; CONES; DATA ACQUISITION; EQUIPMENT; GEOMETRY; IMAGES; PERFORMANCE; RESOLUTION; TESTING; TRAJECTORIES; X RADIATION; X-RAY SOURCES
Descriptors DEC
ELECTROMAGNETIC RADIATION; EVALUATION; IONIZING RADIATIONS; MATHEMATICAL LOGIC; MATHEMATICS; RADIATION SOURCES; RADIATIONS; SIMULATION

Optional Information

Copyright
Copyright (c) 2006 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.