Published January 2004 | Version v1
Journal article

Determination of the Isotope Ratio for Metal Samples Using a Laser Ablation/Ionization Time-of-flight Mass Spectrometry

  • 1. Korea Atomic Energy Research Institute, Daejeon (Korea, Republic of)
  • 2. Chungbuk National University, Jeonju (Korea, Republic of)

Description

The laser ablation/ionization time-of-flight mass spectrometry is applied to the isotopic analysis of solid samples using a home-made instrument. The technique is convenient for solid sample analysis due to the onestep process of vaporization and ionization of the samples. The analyzed samples were lead, cadmium, molybdenum, and ytterbium. To optimize the analytical conditions of the technique, several parameters, such as laser energy, laser wavelength, size of the laser beam on the samples surface, and high voltages applied on the ion source electrodes were varied. Low energy of laser light was necessary to obtain the optimal mass resolution of spectra. The 532 nm light generated mass spectra with the higher signal-to-noise ratio compared with the 355 nm light. The best mass resolution obtained in the present study is ∼1,500 for the ytterbium

Additional details

Publishing Information

Journal Title
Bulletin of the Korean Chemical Society
Journal Volume
25
Journal Issue
1
Series
21 refs, 6 figs, 1 tab
Journal Page Range
p. 101-105
ISSN
0253-2964

INIS

Country of Publication
Korea, Republic of
Country of Input or Organization
Korea, Republic of
INIS RN
42084858
Subject category
S37: INORGANIC, ORGANIC, PHYSICAL AND ANALYTICAL CHEMISTRY;
Descriptors DEI
ABLATION; EVAPORATION; ION SOURCES; IONIZATION; ISOTOPES; MASS RESOLUTION; MASS SPECTROSCOPY; SIGNAL-TO-NOISE RATIO
Descriptors DEC
DIMENSIONLESS NUMBERS; PHASE TRANSFORMATIONS; RESOLUTION; SPECTROSCOPY