Application of avalanche photodiode for soft X-ray pulse-height analyses in the Ht-7 tokamak
Description
An avalanche photodiode (APD) has been used as soft X-ray energy pulse-height analysis system for the measurement of the electron temperature on the HT-7 tokamak. The experimental results obtained with the APD with its inferior energy resolution show a little difference compared to the conventional high energy-resolution Si (Li) detector. Both numerical analysis and experimental results prove that the APD is good enough for application of the electron temperature measurement in tokamaks
Additional details
Identifiers
- PII
- S0168900202005132;
Publishing Information
- Journal Title
- Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
- Journal Volume
- 488
- Journal Issue
- 3
- Journal Page Range
- p. 566-571
- ISSN
- 0168-9002
- CODEN
- NIMAER
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 34008309
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- ELECTRON TEMPERATURE; ENERGY RESOLUTION; HT-7 TOKAMAK; LI-DRIFTED SI DETECTORS; PHOTODIODES; PULSE ANALYZERS; X-RAY EMISSION ANALYSIS
- Descriptors DEC
- CHEMICAL ANALYSIS; CLOSED PLASMA DEVICES; ELECTRONIC EQUIPMENT; EQUIPMENT; LI-DRIFTED DETECTORS; MEASURING INSTRUMENTS; NONDESTRUCTIVE ANALYSIS; RADIATION DETECTORS; RESOLUTION; SEMICONDUCTOR DETECTORS; SEMICONDUCTOR DEVICES; SEMICONDUCTOR DIODES; SI SEMICONDUCTOR DETECTORS; THERMONUCLEAR DEVICES; TOKAMAK DEVICES
Optional Information
- Copyright
- Copyright (c) 2002 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.