The measurement methods and instrumentation used in the realization and transfer of the International Practical Temperature Scale (IPTS-68) above the temperature of freezing gold are described. The determination of the ratios of spectral radiance of tungsten-strip lamps to a gold-point blackbody at a wavelength of 654.6 nm is detailed. The response linearity, spectral responsivity, scattering error, and polarization properties of the instrumentation are described. The analysis of the sources of error and estimates of uncertainty are presented. The assigned uncertainties (three standard deviations) in radiance temperature range from + or - 2 K at 2573 K to + or - 0.5 K at 1073 K

Additional details

Publishing Information

Imprint Title
Noncontact Temperature Measurement
Journal Page Range
vp.
Report number
N--88-23895

Conference

Title
Noncontact temperature measurement conference.
Dates
30 Apr - 1 May 1987.
Place
Washington, DC (USA).

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
20080205
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
ACCURACY; CALIBRATION STANDARDS; RADIOMETERS; US NBS
Descriptors DEC
MEASURING INSTRUMENTS; NATIONAL ORGANIZATIONS; RADIATION DETECTORS; US ORGANIZATIONS