Published March 1988
| Version v1
Report
NBS scale of radiance temperature
Creators
Description
The measurement methods and instrumentation used in the realization and transfer of the International Practical Temperature Scale (IPTS-68) above the temperature of freezing gold are described. The determination of the ratios of spectral radiance of tungsten-strip lamps to a gold-point blackbody at a wavelength of 654.6 nm is detailed. The response linearity, spectral responsivity, scattering error, and polarization properties of the instrumentation are described. The analysis of the sources of error and estimates of uncertainty are presented. The assigned uncertainties (three standard deviations) in radiance temperature range from + or - 2 K at 2573 K to + or - 0.5 K at 1073 K
Additional details
Publishing Information
- Imprint Title
- Noncontact Temperature Measurement
- Journal Page Range
- vp.
- Report number
- N--88-23895
Conference
- Title
- Noncontact temperature measurement conference.
- Dates
- 30 Apr - 1 May 1987.
- Place
- Washington, DC (USA).
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 20080205
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- ACCURACY; CALIBRATION STANDARDS; RADIOMETERS; US NBS
- Descriptors DEC
- MEASURING INSTRUMENTS; NATIONAL ORGANIZATIONS; RADIATION DETECTORS; US ORGANIZATIONS