Analysis of background γ-ray measured with shielded HPGe detector

  • 1. Seoul National Univ., Seoul (Korea, Republic of)

Description

The background γ-ray was measured with shielded HPGe detector. Lead bricks with 10 cm thickness and copper plates with 3 mm thickness were used for the shield. The shield dimension was 1000x1400x1150 mm3. The background was measured with and without the shield as well as with the copper plates inside of the shield. Also, the Χ-rays and γ-rays spectra for standard sources, 241Am, 133Ba, 137Cs, 60Co, 152Eu, were measured. The γ-ray spectroscopy systems was composed of closed-ended coaxial type HPGe detector, Linear AMP, ADC, and AIM. The relative detection efficiency of the HPGe detector is 18% and the energy resolution is 1.8 keV for the 1.3 MeV γ-ray peak from 60Co

Part of:
Proceedings of the KNS spring meeting

Additional details

Publishing Information

Publisher
KNS
Imprint Place
Taejon (Korea, Republic of)
Imprint Title
Proceedings of the KNS spring meeting
Imprint Pagination
[1 CD-ROM]
Journal Page Range
[2 p.]

Conference

Title
2005 spring meeting of the KNS
Dates
26-27 May 2005
Place
Jeju (Korea, Republic of)

INIS

Country of Publication
Korea, Republic of
Country of Input or Organization
Korea, Republic of
INIS RN
37041779
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Resource subtype / Literary indicator
Conference, Non-conventional Literature
Descriptors DEI
BACKGROUND RADIATION; ENERGY RESOLUTION; GAMMA RADIATION; GAMMA SPECTRA; HIGH-PURITY GE DETECTORS; SHIELDS
Descriptors DEC
ELECTROMAGNETIC RADIATION; GE SEMICONDUCTOR DETECTORS; IONIZING RADIATIONS; MEASURING INSTRUMENTS; RADIATION DETECTORS; RADIATIONS; RESOLUTION; SEMICONDUCTOR DETECTORS; SPECTRA

Optional Information

Notes
4 refs, 2 figs, 1 tab