Published 2005
| Version v1
Miscellaneous
Analysis of background γ-ray measured with shielded HPGe detector
Creators
- 1. Seoul National Univ., Seoul (Korea, Republic of)
Description
The background γ-ray was measured with shielded HPGe detector. Lead bricks with 10 cm thickness and copper plates with 3 mm thickness were used for the shield. The shield dimension was 1000x1400x1150 mm3. The background was measured with and without the shield as well as with the copper plates inside of the shield. Also, the Χ-rays and γ-rays spectra for standard sources, 241Am, 133Ba, 137Cs, 60Co, 152Eu, were measured. The γ-ray spectroscopy systems was composed of closed-ended coaxial type HPGe detector, Linear AMP, ADC, and AIM. The relative detection efficiency of the HPGe detector is 18% and the energy resolution is 1.8 keV for the 1.3 MeV γ-ray peak from 60Co
Additional details
Publishing Information
- Publisher
- KNS
- Imprint Place
- Taejon (Korea, Republic of)
- Imprint Title
- Proceedings of the KNS spring meeting
- Imprint Pagination
- [1 CD-ROM]
- Journal Page Range
- [2 p.]
Conference
- Title
- 2005 spring meeting of the KNS
- Dates
- 26-27 May 2005
- Place
- Jeju (Korea, Republic of)
INIS
- Country of Publication
- Korea, Republic of
- Country of Input or Organization
- Korea, Republic of
- INIS RN
- 37041779
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference, Non-conventional Literature
- Descriptors DEI
- BACKGROUND RADIATION; ENERGY RESOLUTION; GAMMA RADIATION; GAMMA SPECTRA; HIGH-PURITY GE DETECTORS; SHIELDS
- Descriptors DEC
- ELECTROMAGNETIC RADIATION; GE SEMICONDUCTOR DETECTORS; IONIZING RADIATIONS; MEASURING INSTRUMENTS; RADIATION DETECTORS; RADIATIONS; RESOLUTION; SEMICONDUCTOR DETECTORS; SPECTRA
Optional Information
- Notes
- 4 refs, 2 figs, 1 tab