Preliminary study of effect of multiple factors of intensity-modulated radiation therapy on dose verification

  • 1. Department of Radiation Oncology, Renji Hospital, School of Medicine, Shanghai Jiao Tong University, Shanghai (China)
  • 2. Department of Oncology, Fourth Affiliated Hospital of Guangxi Medical University, Liuzhou (China)
  • 3. Department of Radiation Oncology, Shanghai Chest Hospital, Shanghai Jiao Tong University, Shanghai (China)
  • 4. Shanghai Institute of Applied Physics, Chinese Academy of Sciences, Shanghai (China)

Description

Objective: To analyze the relationship between planning factors of intensity-modulated radiation therapy (IMRT) and gamma index and investigate the effect of each parameter upon the γ passing rate of IMRT. Methods: Gamma analysis was performed using 3%/3 mm acceptance criteria for 457 IMRT beams with different planning factors. During multi-factor ANOVA analysis of planning factors and gamma passing rate, the control variables primarily included the minimum segment area, minimum number of monitor unit (MU), number of segment, segment conversation, and the spatial resolution in the measured dose distribution. Results: The percentage of pixels with passing γ significantly differed under different minimum segment area, segment conversation and the spatial resolution in the measured dose distribution (all P < 0.05). No significant correlation was observed between the passing rate and the minimum number of MU and the number of segment (P > 0.05). Conclusions: According to the actual situation of the equipment, the minimum segment area should be determined during IMRT planning. Direct machine parameter optimization should be performed. Appropriate resolution of the plane dose images can be chosen according to the minimum detector interval of dose matrix device. (authors)

Additional details

Publishing Information

Journal Title
Chinese Journal of Radiation Oncology
Journal Volume
27
Journal Issue
10
Journal Page Range
p. 933-936
ISSN
1004-4221

INIS

Country of Publication
China
Country of Input or Organization
China
INIS RN
55022121
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
DOSES; EQUIPMENT; RADIATION DOSE DISTRIBUTIONS; RADIOTHERAPY; SPATIAL RESOLUTION; VERIFICATION
Descriptors DEC
MEDICINE; NUCLEAR MEDICINE; RADIOLOGY; RESOLUTION; THERAPY

Optional Information

Notes
2 tabs., 12 refs.; http://dx.doi.org/10.3760/cma.j.issn.1004-4221.2018.10.013