Published October 1986 | Version v1
Journal article

A new method for evaluating weak leakage gamma-ray dose using a 3''φx3'' NaI (Tl) scintillation spectrometer, 5

  • 1. Government Industrial Research Inst., Nagoya (Japan)

Description

Some characteristics of a portable measuring device and the data-processing system were examined. Compared with the manual processing method, the accuracy of the evaluated exposure rate was markedly improved and the measuring time can be reduced to about one-tenth with the same accuracy. At present the reproducibility of this system is about 15 % for 1000 sec accumulating time at the leakage dose rate of 1 μR/h. (author)

Additional details

Additional titles

Subtitle (English)
Characteristics of a portable measuring device and the processing system

Publishing Information

Journal Title
Nagoya Kogyo Gijutsu Shikenjo Hokoku
Journal Volume
35
Journal Issue
9-10
Series
Nagoya Kogyo Gijutsu Shikenjo Hokoku.
Journal Page Range
332-345
ISSN
0027-7614
CODEN
NKGSA