A scatter-corrected list-mode reconstruction and a practical scatter/random approximation technique for dynamic PET imaging
Creators
- 1. Department of Physics and Astronomy, University of British Columbia, Vancouver, BC V6T 1Z1 (Canada)
- 2. Pacific Parkinson's Research Centre, Vancouver, BC V6T 1Z1 (Canada)
Description
We describe an ordinary Poisson list-mode expectation maximization (OP-LMEM) algorithm with a sinogram-based scatter correction method based on the single scatter simulation (SSS) technique and a random correction method based on the variance-reduced delayed-coincidence technique. We also describe a practical approximate scatter and random-estimation approach for dynamic PET studies based on a time-averaged scatter and random estimate followed by scaling according to the global numbers of true coincidences and randoms for each temporal frame. The quantitative accuracy achieved using OP-LMEM was compared to that obtained using the histogram-mode 3D ordinary Poisson ordered subset expectation maximization (3D-OP) algorithm with similar scatter and random correction methods, and they showed excellent agreement. The accuracy of the approximated scatter and random estimates was tested by comparing time activity curves (TACs) as well as the spatial scatter distribution from dynamic non-human primate studies obtained from the conventional (frame-based) approach and those obtained from the approximate approach. An excellent agreement was found, and the time required for the calculation of scatter and random estimates in the dynamic studies became much less dependent on the number of frames (we achieved a nearly four times faster performance on the scatter and random estimates by applying the proposed method). The precision of the scatter fraction was also demonstrated for the conventional and the approximate approach using phantom studies
Additional details
Identifiers
- DOI
- 10.1088/0031-9155/52/8/004;
- PII
- S0031-9155(07)35645-5;
Publishing Information
- Journal Title
- Physics in Medicine and Biology
- Journal Volume
- 52
- Journal Issue
- 8
- Journal Page Range
- p. 2089-2106
- ISSN
- 0031-9155
- CODEN
- PHMBA7
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 38072564
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ACCURACY; ALGORITHMS; APPROXIMATIONS; CORRECTIONS; PHANTOMS; POSITRON COMPUTED TOMOGRAPHY
- Descriptors DEC
- CALCULATION METHODS; COMPUTERIZED TOMOGRAPHY; DIAGNOSTIC TECHNIQUES; EMISSION COMPUTED TOMOGRAPHY; MATHEMATICAL LOGIC; MOCKUP; STRUCTURAL MODELS; TOMOGRAPHY