Published April 1987 | Version v1
Journal article

A comparison of isotope dilution mass spectrometric methods for the assay of copper in copper ore reference materials

  • 1. National Bureau of Standards, Gaithersburg, MD (USA)

Description

The trend toward lower assays in the repeated determination of copper in copper ore, Standard Reference Material 332, have resulted in the withdrawal of this reference material due to material instability. Recent analyses by both isotope dilution thermal ionisation mass spectrometry (TIMS) and isotope dilution inductively coupled plasma mass spectrometry (ICP-MS) have confirmed this change in copper assay. In this study ICP-MS was shown to be an excellent method for the determination of the concentration of copper in copper ore and therefore may be capable of relatively high precision for a number of elements in inorganic matrices. Two experiments were designed to study the relationship between the high loss of volatile components on drying and the corresponding low copper assay. The slow oxidation of CuS to CuSO4 was confirmed. The accelerated moisture study indicated that exposure to changes in temperature and relative humidity did not cause this oxidation. Microwave drying provided additional information about moisture associated with the samples that had been oxidised. (author)

Additional details

Publishing Information

Journal Title
Analyst
Journal Volume
112
Journal Issue
4
Series
Analyst.
Journal Page Range
441-444
ISSN
0003-2654
CODEN
ANALA

Conference

Title
7. SAC international conference on analytical chemistry.
Acronym
SAC 86
Dates
20-26 Jul 1986.
Place
Bristol (UK).