The 1997 determination of the Australian standards of exposure and absorbed dose at 60Co
Creators
- 1. Australian Radiation Laboratory, Yallambie, VIC (Australia)
- 2. Australian Nuclear Science and Technology Organisation, Lucas Heights, NSW (Australia)
Description
The arrangements for the maintenance of the Australian standards for 60Co are described in detail. The primary standards are a graphite cavity chamber for exposure/air kerma and a graphite calorimeter for absorbed dose. These secondary standards are described and their responses in corresponding 90Sr reference sources are reported. Accurate ratios between the Australian Radiation Laboratory (ARL) and Australian Nuclear Science and Technology (ANSTO) 90Sr reference sources are derived for use in future calibrations. The value of 28.8 years for the half-life of 90Sr is confirmed. The usefulness of 90Sr reference source measurements in quality assurance is discussed. The charge sensitivity and linearity of the ANSTO electrometers are reported by two different methods and are compared with previous results. Calibration factors for all the secondary standard ionization chambers are given, in terms of exposure, air kerma and absorbed dose to water. Calibration factors are also given for most of the chambers in terms of absorbed dose to graphite. The methods of deriving the calibration factors are explained in detail, including all the corrections applied to both the primary and secondary standard measurements. Three alternative methods of deriving the absorbed dose to water calibration factors are compared. The reported calibration factors are compared with previous results. Changes in the Australian units of exposure, air kerma and absorbed dose to graphite and water are derived from changes in the corresponding calibration factors. The Australian units of exposure and air kerma have not changed significantly since 1990. The Australian unit of absorbed dose to graphite is now 1.1 % smaller than in 1993 and 1.3 % smaller than in 1990. The Australian unit of absorbed dose to water is now 1.4 % smaller than in 1993, but is only 0.9 % smaller than in 1990. Comparisons of the Australian standards of exposure/air kerma and absorbed dose with those of the Bureau International des Poids et Mesures (BIPM) were performed in 1997 and show that the Australian units fall within the range of those of other countries. (authors)
Availability note (English)
Available from INIS in electronic formFiles
30003756.pdf
Files
(1.2 MB)
| Name | Size | Download all |
|---|---|---|
|
md5:2d7c19157ee7f78a2f4cd0e6b406c88c
|
1.2 MB | Preview Download |
Additional details
Publishing Information
- Imprint Pagination
- 65 p.
- Report number
- ARL/TR--126
INIS
- Country of Publication
- Australia
- Country of Input or Organization
- Australia
- INIS RN
- 30003756
- Subject category
- S07: ISOTOPES AND RADIATION SOURCES;
- Resource subtype / Literary indicator
- Numerical Data
- Descriptors DEI
- CALIBRATION STANDARDS; COBALT 60; COMPARATIVE EVALUATIONS; ELECTROMETERS; EVALUATED DATA; INTERLABORATORY COMPARISONS; IONIZATION CHAMBERS; RADIATION DOSES; STRONTIUM 90
- Descriptors DEC
- BETA DECAY RADIOISOTOPES; BETA-MINUS DECAY RADIOISOTOPES; COBALT ISOTOPES; DATA; ELECTRIC MEASURING INSTRUMENTS; ELECTRICAL EQUIPMENT; EQUIPMENT; EVALUATION; EVEN-EVEN NUCLEI; INFORMATION; INTERMEDIATE MASS NUCLEI; INTERNAL CONVERSION RADIOISOTOPES; ISOMERIC TRANSITION ISOTOPES; ISOTOPES; MEASURING INSTRUMENTS; MINUTES LIVING RADIOISOTOPES; NUCLEI; NUMERICAL DATA; ODD-ODD NUCLEI; RADIATION DETECTORS; RADIOISOTOPES; STANDARDS; STRONTIUM ISOTOPES; YEARS LIVING RADIOISOTOPES
Optional Information
- Notes
- 31refs., 19 tabs., 8 figs.