Published March 1993
| Version v1
Journal article
Is it possible to verify directly a proton-treatment plan using positron emission tomography?
- 1. UCL-Cliniques Universitaires St-Luc, Brussels (Belgium)
- 2. Positron Tomography Laboratory, UCL, Louvain-la-Neuve (Belgium)
Description
A PET camera is used to visualize the positron activity induced during protonbeam therapy in order to verify directly the proton-treatment plans. The positron emitters created are predominantly the 15O and 11C, whose total activity amounts to 12 MBq after an irradiation with 85 MeV protons, delivering 3 Gy in a volume of approximately 300 cm3. Although this method is a useful verification of patient setup, care must be taken when deriving dose distributions from activity distributions. Correlation between both quantities is difficult, moreover at the last millimeters of their range, protons will no longer activate tissue. Due to the short half-lives the PET camera must be located close to the treatment facility. (author) 17 refs
Additional details
Publishing Information
- Journal Title
- Radiotherapy and Oncology
- Journal Volume
- 26
- Journal Issue
- 3
- Journal Page Range
- p. 275-277.
- ISSN
- 0167-8140
- CODEN
- RAONDT
INIS
- Country of Publication
- Ireland
- Country of Input or Organization
- Ireland
- INIS RN
- 25003000
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- NUCLEAR MEDICINE; POSITRON BEAMS; POSITRON CAMERAS; POSITRON COMPUTED TOMOGRAPHY; POSITRON DETECTION; PROTON BEAMS; RADIOTHERAPY
- Descriptors DEC
- BEAMS; CAMERAS; CHARGED PARTICLE DETECTION; COMPUTERIZED TOMOGRAPHY; DETECTION; EMISSION COMPUTED TOMOGRAPHY; GAMMA CAMERAS; LEPTON BEAMS; MEDICINE; NUCLEON BEAMS; PARTICLE BEAMS; RADIATION DETECTION; THERAPY; TOMOGRAPHY