Published December 1978 | Version v1
Journal article

Studies on the quality control of insulin radioimmunoassay kit pt. 1

  • 1. Korea Atomic Energy Research Inst., Seoul (Republic of Korea)

Description

A typical abnormal standard dose response curve of nearly convex form often encountered in insulin radioimmunoassay (IRIA) has been analyzed by varying incubation conditions. The cause of such abnormality has been turned out to be the results of an incomplete equilibrium between the two reactants. By careful control of the temperature of serum sample an immediate cause of sharp deviation of B/F value of the sample tube from the measurable range in the standard curve has also been investigated. The two main troubles have been proven to be stemmed from incubation conditions. Incubation at 40C for 48 hrs is emphasized for IRIA. (author)

Additional details

Additional titles

Subtitle (English)
Pitfalls in radioimmunoassay of insulin

Publishing Information

Journal Title
J. Korean Nucl. Soc.
Journal Volume
10
Journal Issue
4
Series
J. Korean Nucl. Soc.
Journal Page Range
223-229
ISSN
0372-7327

INIS

Country of Publication
Korea, Republic of
Country of Input or Organization
Korea, Republic of
INIS RN
11517453
Subject category
S60: APPLIED LIFE SCIENCES;
Descriptors DEI
ANTIBODIES; FAILURE MODE ANALYSIS; INCUBATION; INSULIN; QUALITY CONTROL; RADIOIMMUNOASSAY; STANDARDIZATION
Descriptors DEC
CONTROL; HORMONES; ISOTOPE APPLICATIONS; PEPTIDE HORMONES; SYSTEM FAILURE ANALYSIS; SYSTEMS ANALYSIS; TRACER TECHNIQUES