To test the hypothesis that quality clinical audits improve compliance with the procedures in computed tomography (CT) scanning. This retrospective study was conducted in two hospitals, based on 6950 examinations and four procedures, focusing on the acquisition length in lumbar spine CT, the default tube current applied in abdominal un-enhanced CT, the tube potential selection for portal phase abdominal CT and the use of a specific ''paediatric brain CT'' procedure. The first clinical audit reported compliance with these procedures. After presenting the results to the stakeholders, a second audit was conducted to measure the impact of this information on compliance and was repeated the next year. Comparisons of proportions were performed using the Chi-square Pearson test. Depending on the procedure, the compliance rate ranged from 27 to 88 % during the first audit. After presentation of the audit results to the stakeholders, the compliance rate ranged from 68 to 93 % and was significantly improved for all procedures (P ranging from <0.001 to 0.031) in both hospitals and remained unchanged during the third audit (P ranging from 0.114 to 0.999). Quality improvement through repeated compliance audits with CT procedures durably improves this compliance. (orig.)

Availability note (English)

Available from: http://dx.doi.org/10.1007/s00330-015-4100-x

Additional details

Identifiers

Publishing Information

Journal Title
European Radiology
Journal Volume
26
Journal Issue
8
Journal Page Range
p. 2663-2668
ISSN
0938-7994
CODEN
EURAE3