  • 1. TRIUMF, Vancouver, BC (Canada)
  • 2. Deutsches Krebsforschungszentrum, Heidelberg (Germany). Inst. fuer Radiologie und Pathophysiologie

Description

The increased scatter fraction seen in positron volume imaging when compared with conventional slice positron emission tomography has created a need for better characterization and correction of scattered gamma rays for positron imaging. An analytical simulation of single scattered gamma rays and an extension to estimate multiple scattered rays have been verified using a Monte Carlo simulation. The simulation was first verified using measured tomograph data. The analytical simulation has been used to study some cases of interest for scatter correction. A scatter correction method which uses the analytical simulation has been proposed. (Author) (13 refs., 6 figs.)