A new scintillator-based fast-ion loss detector (FILD) installed near the outer midplane of the plasma has been commissioned on DIII-D. This detector successfully measures coherent fast ion losses produced by fast-ion driven instabilities (≤500 kHz). Combined with the first FILD at ∼45° below the outer midplane [R. K. Fisher, et al., Rev. Sci. Instrum. 81, 10D307 (2010)], the two-detector system measures poloidal variation of losses. The phase space sensitivity of the new detector (gyroradius rL∼[1.5–8] cm and pitch angle α∼[35°–85°]) is calibrated using neutral beam first orbit loss measurements. Since fast ion losses are localized poloidally, having two FILDs at different poloidal locations allows for the study of losses over a wider range of plasma shapes and types of loss orbits.
Additional details
Identifiers
- DOI
- 10.1063/1.4732063;
Publishing Information
- Journal Title
- Review of Scientific Instruments
- Journal Volume
- 83
- Journal Issue
- 10
- Journal Page Range
- p. 10D707-10D707.3
- ISSN
- 0034-6748
- CODEN
- RSINAK
INIS
- Country of Publication
- United States
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 44052256
- Subject category
- S70: PLASMA PHYSICS AND FUSION TECHNOLOGY; S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- BEAMS; DOUBLET-3 DEVICE; INCLINATION; IONS; LOSSES; PHASE SPACE; PLASMA; PLASMA BEAM INJECTION; PLASMA CONFINEMENT; PLASMA INSTABILITY; SCINTILLATION COUNTERS; SENSITIVITY; THERMONUCLEAR REACTORS
- Descriptors DEC
- BEAM INJECTION; CHARGED PARTICLES; CLOSED PLASMA DEVICES; CONFINEMENT; INSTABILITY; MATHEMATICAL SPACE; MEASURING INSTRUMENTS; RADIATION DETECTORS; SPACE; THERMONUCLEAR DEVICES; TOKAMAK DEVICES
Optional Information
- Notes
- (c) 2012 American Institute of Physics