Published March 2016 | Version v1
Journal article

18F primary standard at ENEA-INMRI by three absolute techniques and calibration of a well-type IG11 ionization chamber

Description

A new 18F primary standardization carried out at ENEA-INMRI by three different absolute techniques, i.e. 4πγNaI(Tl)γ high-efficiency counting, TDCR and 4πβ(LS)-γ[NaI(Tl)] coincidence counting method, allowed the calibration of a fixed well-reentrant IG11 ionization chamber (IC), with an uncertainty lower than 1%, and to check the calibration factor of a portable well-type IC NPL-CRC model, previously calibrated. By the new standard the ENEA-INMRI was linked to the BIPM International Reference System (SIR) through the BIPM SIR Transfer Instrument (SIRTI). - Highlights: • New 18F primary standardization performed at ENEA-INMRI. • Absolute methods: 4πγ integral counting, TDCR and 4πβ(LS)-γ coincidence counting. • Two Ionization Chambers calibration: NPL-CRC Model and IG11. • New software code MAC4.cc for coincidence analysis in a beta(LS)-gamma system. • Link of the ENEA-INMRI to the BIPM SIR through the SIRTI instrument.

Availability note (English)

Available from http://dx.doi.org/10.1016/j.apradiso.2015.12.055

Additional details

Identifiers

DOI
10.1016/j.apradiso.2015.12.055;
PII
S0969-8043(15)30396-1;

Publishing Information

Journal Title
Applied Radiation and Isotopes
Journal Volume
109
Journal Page Range
p. 410-413
ISSN
0969-8043
CODEN
ARISEF

Conference

Title
20. international conference on radionuclide metrology and its applications
Dates
8-11 Jun 2015
Place
Vienna (Austria)

Optional Information

Copyright
Copyright (c) 2016 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.