Published 1983
| Version v1
Miscellaneous
The Q3D magnetic spectrograph
Description
During the year 1983 several changes have been done at the Q3D system, especially to broaden the range of possible reactions for mass measurements: A new bending magnet in the beam line deflects ions with B x rho up to 15 kG x m; a new focal plane detector with better dE and E resolution allows the separation of ions with masses up to A approx.= 40 together with strong background suppression, and an electronically controlled gas flow system shows excellent long-time stability, thus allowing the measurement of reactions with very low cross sections. (orig.)
Availability note (English)
Available from Max-Planck-Institut fuer Kernphysik, Heidelberg (Germany, F.R.).Additional details
Additional titles
- Original title (German)
- Der Q3D Magnetspektrograph
Publishing Information
- Imprint Title
- Max-Planck-Institut fuer Kernphysik. Annual report 1983
- Imprint Pagination
- 223 p.
- Journal Page Range
- p. 26-27.
INIS
- Country of Publication
- Germany
- Country of Input or Organization
- Germany
- INIS RN
- 16003299
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Non-conventional Literature
- Descriptors DEI
- BACKGROUND NOISE; BEAM BENDING MAGNETS; ELECTRONIC GUIDANCE; ENERGY RESOLUTION; HEAVY ION SPECTROMETERS; ION BEAMS; MAGNETIC DIPOLES; MAGNETIC SPECTROMETERS; QUADRUPOLES; STABILITY
- Descriptors DEC
- BEAMS; CONTROL SYSTEMS; DIPOLES; EQUIPMENT; MAGNETS; MEASURING INSTRUMENTS; MULTIPOLES; NOISE; RESOLUTION; SPECTROMETERS
Optional Information
- Notes
- Imprint:Max-Planck-Institut fuer Kernphysik. Jahresbericht 1983.