Towards the definition of the optimum height of voluminous cylindrical sources in gamma-ray spectroscopy
Creators
- 1. Physics Department, Faculty of Mining, Bor (Yugoslavia)
- 2. Faculty of Physics, Belgrade (Yugoslavia)
- 3. Institute of Nuclear Sciences 'Vinca', Belgrade (Yugoslavia)
Description
We show that for cylindrical voluminous sources in gamma-ray spectroscopy the relative error of the spectral intensity of a low-energy spectral line, as the height of the source increases, saturates faster than the spectral intensity itself. This significantly reduces the quantity of the required sample material as compared to that based on the usual considerations of the spectral intensity only. We also demonstrate that there exists a weakly pronounced height of the source (the source size) which minimizes the relative error of the intensity of a given low-energy spectral line. Due to the smallness of the effect the benefit of introducing the concept of the truly optimum size of the source is more on a psychological side, one should feel easier using an optimum quantity of the source material than merely a sufficient one, as is otherwise the case
Additional details
Identifiers
- PII
- S0969804398001729;
Publishing Information
- Journal Title
- Applied Radiation and Isotopes
- Journal Volume
- 50
- Journal Issue
- 6
- Journal Page Range
- p. 999-1001
- ISSN
- 0969-8043
- CODEN
- ARISEF
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 34015165
- Subject category
- S07: ISOTOPES AND RADIATION SOURCES;
- Descriptors DEI
- CYLINDRICAL CONFIGURATION; ERRORS; GAMMA SOURCES; GAMMA SPECTRA; HEIGHT; OPTIMIZATION; VOLUME
- Descriptors DEC
- CONFIGURATION; DIMENSIONS; RADIATION SOURCES; SPECTRA