Published November 1978
| Version v1
Journal article
Patient dosage in computed tomography
Description
The maximum surface dosage in most clinical CT scans seems to range from 2 to 10 rads/study but much larger dose per study values seem possible with both rotate-translate and rotary geometry designs. The CT scanner type in itself does not significantly reduce doses. Secondary radiation dose values were measured for critical organs and indicate that dosage from secondary radiations may be reduced significantly by external shielding. Dose values in the vicinity of most CT scanners are typically 1 to 2 mrad/scan at 1 meter at the parameters of a typical clinical scan
Additional details
Publishing Information
- Journal Title
- Radiology
- Journal Volume
- 129
- Journal Issue
- 2
- Series
- Radiology.
- Journal Page Range
- 457-463
Conference
- Title
- 62. scientific assembly and annual meeting of the Radiological Society of North America.
- Dates
- 14 Nov 1976.
- Place
- Chicago, IL, USA.
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 10445019
- Subject category
- S63: RADIATION, THERMAL, AND OTHER ENVIRONMENTAL POLLUTANT EFFECTS ON LIVING ORGANISMS AND BIOLOGICAL MATERIALS; S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- CAT SCANNING; PATIENTS; RADIATION DOSES; SCATTERING; SHIELDING; X RADIATION; X-RAY DOSIMETRY
- Descriptors DEC
- DIAGNOSTIC TECHNIQUES; DOSIMETRY; ELECTROMAGNETIC RADIATION; IONIZING RADIATIONS; PHOTON TRANSMISSION SCANNING; RADIATIONS