Precision calibration of calorimeter electronics in the D0 liquid argon/uranium particle detector

Description

The ability to cross calibrate thousands of channels of detector electronics is of prime importance. This paper will describe the system used to deliver and distribute a 300 nanosecond pulse across 50,000 channels of electronics with better than 0.25% difference between channels from a location more than 200 feet away. The system is used for both cross calibration and functionality checking, (i.e., missing channels). Design of a fixed width pulse generator of high stability is presented as a key ingredient in the system's overall performance. In addition, the design of a controlled impedance distribution system is discussed. 2 refs

Availability note (English)

MF available from INIS under the Report Number; OSTI as DE92007203; NTIS; INIS; US Govt. Printing Office Dep.

Files

Restricted

The record is publicly accessible, but files are restricted to users with access.

Additional details

Publishing Information

Imprint Pagination
6 p.
Report number
FNAL/C--91/344

Conference

Title
IEEE nuclear science symposium.
Dates
5-9 Nov 1991.
Place
Santa Fe, NM (United States).

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
23040289
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S43: PARTICLE ACCELERATORS;
Resource subtype / Literary indicator
Conference
Descriptors DEI
CALIBRATION; DESIGN; ELECTROMAGNETIC FILTERS; PREAMPLIFIERS; PULSE GENERATORS; PULSE TECHNIQUES; SHOWER COUNTERS; TRANSFORMERS
Descriptors DEC
AMPLIFIERS; ELECTRICAL EQUIPMENT; ELECTRONIC EQUIPMENT; EQUIPMENT; FILTERS; FUNCTION GENERATORS; MEASURING INSTRUMENTS; RADIATION DETECTORS

Optional Information

Contract/Grant/Project number
Contract AC02-76CH03000
Funding organization
USDOE, Washington, DC (United States).
Secondary number(s)
CONF-911106--70.