Mammography calibration qualities establishment in a Mo- Mo clinical system
- 1. Nuclear and Energy Research Institute, University of São Paulo, Av. Prof. Lineu Prestes, 2242, São Paulo, SP (Brazil)
Description
In this study the mammography calibration qualities were established in a clinical mammography system. The objective is to provide the IPEN instruments calibration laboratory with both mammography calibration methods (using a clinical and an industrial system). The results showed a good behavior of mammography equipment, in terms of kVp, PPV and exposure time. The additional filtration of molybdenum is adequate, air-kerma rates were determined and spectra were obtained. (paper)
Availability note (English)
Available from http://dx.doi.org/10.1088/1742-6596/733/1/012086Additional details
Identifiers
Publishing Information
- Journal Title
- Journal of Physics. Conference Series (Online)
- Journal Volume
- 733
- Journal Issue
- 1
- Journal Page Range
- [5 p.]
- ISSN
- 1742-6596
Conference
- Title
- 8. Brazilian congress on metrology
- Acronym
- Metrologia 2015
- Dates
- 29 Nov - 2 Dec 2015
- Place
- Bento Goncalves, RS (Brazil)
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 48100485
- Subject category
- S07: ISOTOPES AND RADIATION SOURCES; S60: APPLIED LIFE SCIENCES; S71: CLASSICAL AND QUANTUM MECHANICS, GENERAL PHYSICS;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- CALIBRATION; FILTRATION; KERMA; MAMMARY GLANDS; MOLYBDENUM; SPECTRA
- Descriptors DEC
- BODY; ELEMENTS; GLANDS; METALS; ORGANS; REFRACTORY METALS; SEPARATION PROCESSES; TRANSITION ELEMENTS