Boron-coated straws as a replacement for 3He-based neutron detectors
Creators
- 1. Proportional Technologies, Inc., 8022 El Rio Street, Houston, TX 77054 (United States)
Description
US and international government efforts to equip major seaports with large area neutron detectors, aimed to intercept the smuggling of nuclear materials, have precipitated a critical shortage of 3He gas. It is estimated that the annual demand of 3He for US security applications alone is more than the worldwide supply. This is strongly limiting the prospects of neutron science, safeguards, and other applications that rely heavily on 3He-based detectors. Clearly, alternate neutron detection technologies that can support large sensitive areas, and have low gamma sensitivity and low cost must be developed. We propose a low-cost technology based on long copper tubes (straws), coated on the inside with a thin layer of 10B-enriched boron carbide (10B4C). In addition to the high abundance of boron on Earth and low cost of 10B enrichment, the boron-coated straw (BCS) detector offers distinct advantages over conventional 3He-based detectors, and alternate technologies such as 10BF3 tubes and 10B-coated rigid tubes. These include better distribution inside moderator assemblies, many-times faster electronic signals, no pressurization, improved gamma-ray rejection, no toxic or flammable gases, and ease of serviceability. We present the performance of BCS detectors dispersed in a solid plastic moderator to address the need for portal monitoring. The design adopts the outer dimensions of currently deployed 3He-based monitors, but takes advantage of the small BCS diameter to achieve a more uniform distribution of neutron converter throughout the moderating material. We show that approximately 63 BCS detectors, each 205 cm long, distributed inside the moderator, can match or exceed the detection efficiency of typical monitors fitted with a 5 cm diameter 3He tube, 187 cm long, pressurized to 3 atm.
Availability note (English)
Available from http://dx.doi.org/10.1016/j.nima.2010.09.011Additional details
Identifiers
- DOI
- 10.1016/j.nima.2010.09.011;
- PII
- S0168-9002(10)01973-X;
Publishing Information
- Journal Title
- Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
- Journal Volume
- 652
- Journal Issue
- 1
- Journal Page Range
- p. 359-363
- ISSN
- 0168-9002
- CODEN
- NIMAER
Conference
- Title
- 12. Symposium on radiation measurements and applications (SORMA)
- Dates
- 24-28 May 2010
- Place
- Ann Arbor, MI (United States)
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 44001844
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- BORON; BORON 10; BORON CARBIDES; COPPER; DESIGN; ENRICHMENT; GAMMA RADIATION; HELIUM 3; NEUTRON CONVERTERS; NEUTRON DETECTORS; PRESSURIZATION; RADIATION MONITORING; SAFEGUARDS; SECURITY; STRAW; SURFACE COATING; THIN FILMS; TOXICITY
- Descriptors DEC
- BORON COMPOUNDS; BORON ISOTOPES; CARBIDES; CARBON COMPOUNDS; DEPOSITION; ELECTROMAGNETIC RADIATION; ELEMENTS; EVEN-ODD NUCLEI; FILMS; HELIUM ISOTOPES; IONIZING RADIATIONS; ISOTOPES; LIGHT NUCLEI; MEASURING INSTRUMENTS; METALS; MONITORING; NUCLEI; ODD-ODD NUCLEI; RADIATION DETECTORS; RADIATIONS; SEMIMETALS; STABLE ISOTOPES; TRANSITION ELEMENTS
Optional Information
- Copyright
- Copyright (c) 2010 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.