Published 1980
| Version v1
Report
Restricted
Method for calibration of plutonium NDA
Description
Calibration materials characterized by calorimetric assay can be a practical alternative to synthetic standards for the calibration of plutonium nondestructive assay. Calorimetric assay is an effective measurement system for the characterization because: it can give an absolute assay from first principles when the isotopic composition is known, it is insensitive to most matrix effects, and its traceability to international measurement systems has been demonstrated
Availability note (English)
MF available from INIS under the Report Number; Available from NTIS., PC A02/MF A01.
Files
Additional details
Publishing Information
- Imprint Pagination
- 5 p.
- Report number
- MLM--2755(OP)
Conference
- Title
- International meeting on monitoring of Pu-contaminated waste.
- Dates
- 25 - 28 Sep 1979.
- Place
- Ispra, Italy.
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 11566381
- Subject category
- S98: NUCLEAR DISARMAMENT, SAFEGUARDS AND PHYSICAL PROTECTION; S37: INORGANIC, ORGANIC, PHYSICAL AND ANALYTICAL CHEMISTRY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- CALIBRATION; CALORIMETRY; NONDESTRUCTIVE ANALYSIS; PLUTONIUM; PLUTONIUM 238; PLUTONIUM 239; RADIOACTIVE WASTES
- Descriptors DEC
- ACTINIDE NUCLEI; ACTINIDES; ALPHA DECAY RADIOISOTOPES; CHEMICAL ANALYSIS; ELEMENTS; EVEN-EVEN NUCLEI; EVEN-ODD NUCLEI; HEAVY NUCLEI; ISOTOPES; METALS; NUCLEI; PLUTONIUM ISOTOPES; RADIOACTIVE MATERIALS; RADIOISOTOPES; TRANSURANIUM ELEMENTS; WASTES; YEARS LIVING RADIOISOTOPES
Optional Information
- Secondary number(s)
- CONF-790928--2.