Published July 1987 | Version v1
Journal article

Effect of computed tomography noise and tissue heterogeneity on cerebral blood flow determination by xenon-enhanced computed tomography

  • 1. Univ. of Pittsburgh, PA

Description

The errors associated with derivation of cerebral blood flow values by the xenon-enhanced computed tomography (CT) method have been evaluated as a function of tissue heterogeneity and CT noise. The results of this study indicate that CT noise introduces large errors in the derived flow value when data for a single, unprocessed voxel are used for this purpose. CT noise increases the derived flow values in a systematic way. Tissue heterogeneity results in a systematic error which lowers the derived flow values. Errors due to both parameters are computed for typical and extreme conditions

Additional details

Publishing Information

Journal Title
Med. Phys.
Journal Volume
14
Journal Issue
4
Series
Med. Phys.
Journal Page Range
557-561
ISSN
0094-2405
CODEN
MPHYA

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
19018292
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
BLOOD FLOW; BRAIN; EMISSION COMPUTED TOMOGRAPHY; ERRORS; XENON ISOTOPES
Descriptors DEC
BODY; CENTRAL NERVOUS SYSTEM; COMPUTERIZED TOMOGRAPHY; NERVOUS SYSTEM; ORGANS; TOMOGRAPHY