Published December 2007 | Version v1
Journal article

Investigation of the response function of the Schottky CdTe detector with guard-ring cathode through Monte Carlo calculation and measurement

  • 1. Tohoku Univ., Cyclotron and Radioisotope Center, Sendai, Miyagi (Japan)
  • 2. Tohoku Univ., Graduate School of Pharmaceutical Sciences, Sendai, Miyagi (Japan)

Description

We investigated the response function of a Schottky CdTe detector with a guard-ring cathode through Monte Carlo calculation using the MCNP code and measurement in order to apply the detector for the analysis of X-and γ-ray spectra in medical radio-diagnostic units for quality assurance and quality control of the systems. The CdTe detector has high potential as an X- and γ-ray spectrometer owing to compactness due to a large photon absorption cross section and high energy resolution. We measured the detector response function for various operating conditions, i.e., bias voltage and shaping time using an 241Am source to find the optimum operating condition, and then measured the detector responses and full energy absorption efficiency for 241Am, 152Eu, 133Ba and finally compared with the calculation. The experimental response functions are well reproduced by the MCNP code except for the part of the low energy tail due to charge trapping effect. (author)

Availability note (English)

Available from http://dx.doi.org/10.12950/rsm2002.6.14

Additional details

Identifiers

Publishing Information

Journal Title
Radiation Safety Management (Online)
Journal Volume
6
Journal Issue
1
Journal Page Range
p. 14-17
ISSN
1884-9520

Optional Information

Notes
8 refs., 5 figs.