10 channel grating polychromator for the measurements of electron cyclotron emission

Description

A 10-channel grating polychromator has been developed and applied for the measurements of the second harmonic electron cyclotron emission from the JIPP T-IIU tokamak plasma. The frequency resolution measured at 150 GHz is about 3 GHz which corresponds to the spatial resolution of 1.8 cm near the center of the plasma column. The grating efficiency is about 80 % for S-polarization around 160 GHz. Its excellent time resolution, Δt ≤ 2 μs, is favorable for studying MHD activities in hot plasmas. It has been found from the measurements in JIPP T-IIU that the grating polychromator is a powerful diagnostic tool for the measurements of electron temperature of the plasma. (author)

Availability note (English)

MF available from INIS under the Report Number.

Files

20031280.pdf

Files (374.2 kB)

Name Size Download all
md5:8000b94889f05f5bbc4883be23779a99
374.2 kB Preview Download

Additional details

Publishing Information

Imprint Pagination
32 p.
Report number
IPPJ--893