Shielding research of minitype reference radiation device based on Monte Carlo simulation
- 1. Institute of Nuclear Physics and Chemistry, CAEP, Mianyang (China)
- 2. Faculty of Optical Engineering, Chongqing University, Chongqing (China)
Description
When applied the minitype reference radiation (MRR) to calibrate gamma ray dosimeters, the shielding structure of the MRR should meet the shielding technique demand, in which the limit dose rate of the outside surface of the MRR must be lower than 5 μSv/h. In order to optimize the shielding design of the MRR, Monte Carlo particle transport code MCNP is employed to simulate and analyze the dose distribution of the outside surface of the MRR. The results show that the dose distribution of each surface of the MRR can be described in detail with MCNP and it is conducive to complete shielding design. The lead-steel composite material is applied to increase the strength and reduce the weight of the MRR. The optimized weight of the MRR is about 271.9 kg. (authors)
Additional details
Identifiers
Publishing Information
- Journal Title
- High Power Laser and Particle Beams
- Journal Volume
- 28
- Journal Issue
- 9
- Journal Page Range
- [8 p.]
- ISSN
- 1001-4322
INIS
- Country of Publication
- China
- Country of Input or Organization
- China
- INIS RN
- 51096316
- Subject category
- S07: ISOTOPES AND RADIATION SOURCES; S61: RADIATION PROTECTION AND DOSIMETRY;
- Descriptors DEI
- COMPOSITE MATERIALS; COMPUTERIZED SIMULATION; DESIGN; DOSE RATES; DOSEMETERS; GAMMA RADIATION; LEAD; MONTE CARLO METHOD; RADIATION DOSE DISTRIBUTIONS; SHIELDING; STEELS
- Descriptors DEC
- ALLOYS; CALCULATION METHODS; CARBON ADDITIONS; ELECTROMAGNETIC RADIATION; ELEMENTS; IONIZING RADIATIONS; IRON ALLOYS; IRON BASE ALLOYS; MATERIALS; MEASURING INSTRUMENTS; METALS; RADIATIONS; SIMULATION; TRANSITION ELEMENT ALLOYS
Optional Information
- Notes
- 8 figs., 3 tabs., 13 refs.; http://dx.doi.org/10.11884/HPLPB201628.160018